Medical Devices

It is imperative that medical devices are kept clean and sterilized for procedures and surgeries. This category has an extensive list of standards relating to specific aspects of medical devices and more.

ANSI/AAMI ST67:2019

Sterilization of health care products - Requirements and guidance for selecting a sterility assurance level (SAL) for products labeled “sterile”

This standard establishes requirements, guidance and a risk framework for identification of acceptable approach(es) for assurance of sterility and for selection of alternative SAL(s) or aseptic processing for health care product that cannot be terminally sterilized to achieve an SAL value of 10 to the -6 

ISO 11139:2018

Sterilization of health care products - Vocabulary of terms used in sterilization and related equipment and process standards

This document defines terms in the field of the sterilization of health care products including related equipment and processes.

ANSI/AAMI ST77:2013 (R2018)

Containment devices for reusable medical device sterilization

This standard covers minimum labeling and performance requirements for rigid sterilization container systems and for instrument organizers.

ISO 10993-7:2008

Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals

ISO 10993-7:2008 specifies allowable limits for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) in individual EO-sterilized medical devices, procedures for the measurement of EO and ECH, and methods for determining compliance so that devices may be released. Additional background, including guidance and a flowchart showing how the standard is applied are also included in informative annexes.

EO-sterilized devices that have no patient contact (e.g., in vitro diagnostic devices) are not covered by ISO 10993-7:2008.

ISO 11135:2014

Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices

ISO 11135:2014 specifies requirements for the development, validation and routine control of an ethylene oxide sterilization process for medical devices in both the industrial and health care facility settings, and it acknowledges the similarities and differences between the two applications.

AAMI TIR74:2016 (R2021)

Change summary for ISO 11135:2014, Sterilization of health care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices.

Provides summary of differences between the new 2014 version and the 2007 version of ISO 11135. Provides end users with a quick reference when evaluating and implementing the 2014 version of ISO 11135 in their facilities.

ISO 13408-2:2018

Aseptic processing of health care products - Part 2: Sterilizing filtration

ISO 13408-2:2018 specifies requirements for sterilizing filtration as part of aseptic processing of health care products conducted in accordance with ISO 13408‑1. It also offers guidance to filter users concerning general requirements for set-up, validation and routine operation of a sterilizing filtration process.

ISO 13408-2:2018 is not applicable to removal of viruses.

Sterilizing filtration is not applicable to fluids that intentionally contain particles larger than the pore size of the filter (e.g. bacterial whole-cell vaccines).

ISO 13408-2:2018 is not applicable to high efficiency particulate air (HEPA) filters.

ISO 13408-2:2018 does not specify requirements for the development, validation and routine control of a process for removing the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease. Specific recommendations have been produced in particular countries for the processing of materials potentially contaminated with these agents.

ISO 13408-5:2006

Aseptic processing of health care products - Part 5: Sterilization in place

ISO 13408-5:2006 specifies the general requirements for sterilization in place (SIP) applied to product contact surfaces of the equipment used in the manufacture of sterile health care products by aseptic processing and offers guidance on qualification, validation, operation and control.

ISO 13408-5:2006 applies to processes where sterilizing agents are delivered to the internal surfaces of equipment that can come in contact with the product.

ISO 14160:2020

Sterilization of health care products - Liquid chemical sterilizing agents for single-use medical devices utilizing animal tissues and their derivatives - Requirements for characterization, development, validation and routine control of a sterilization process for medical devices

This document specifies requirements for the characterization of a liquid chemical sterilizing agent and for the development, validation, process control and monitoring of sterilization by liquid chemical sterilizing agents of single-use medical devices comprising, in whole or in part, materials of animal origin. This document covers the control of risks arising from contamination with bacteria and fungi by application of a liquid chemical sterilization process. Risks associated with other microorganisms can be assessed using other methods (see NOTE 1). This document is not applicable to material of human origin. This document does not describe methods for the validation of the inactivation of viruses and transmissible spongiform encephalopathy (TSE) agents (see NOTE 2 and NOTE 3). This document does not describe methods for validation of the inactivation or elimination of protozoa and parasites. The requirements for validation and routine control described in this document are only applicable to the defined sterilization process of a medical device, which is performed after the manufacturing process, and do not take account of the lethal effects of other bioburden reduction steps (see NOTE 4). This document does not specify tests to establish the effects of any chosen sterilization process upon the fitness for use of the medical device (see NOTE 5). This document does not cover the level of residual sterilizing agent within medical devices (see NOTE 6). Guidance for the characterization of a liquid chemical sterilizing agent and for the development, validation, process control and monitoring of sterilization by liquid chemical sterilizing agents of single-use medical devices comprising, in whole or in part, materials of animal origin is provided in informative Annex A. NOTE 1 The prior application of risk management principles to medical devices utilizing animal tissues, as described in ISO 22442-1 is important. ISO 18362 provides information on control of microbial risks during processing of cell-based health-care products. NOTE 2 Liquid chemical sterilizing agents traditionally employed to sterilize animal tissues in medical devices might not be effective in inactivating the causative agents of TSE such as bovine spongiform encephalopathy (BSE), or scrapie. Satisfactory validation in accordance with this document does not necessarily demonstrate inactivation of infective agents of this type. Risk controls related to sourcing, collection and handling of animal materials are described in ISO 22442-2. NOTE 3 The validation of the inactivation, elimination, or elimination and inactivation of viruses and TSE agents is described in ISO 22442-3. NOTE 4 Manufacturing processes for medical devices containing animal tissues frequently include exposure to chemical agents which can significantly reduce the bioburden on the medical device. Following the manufacturing process, a medical device is exposed to a specified sterilization process. NOTE 5 Such testing is a crucial part of the design and development of a medical device. NOTE 6 ISO 10993-17 specifies a method to establish allowable limits for residues of sterilizing agents. NOTE 7 Standards for quality management systems (see ISO 13485) can be used in the control of all stages of manufacture including the sterilization process.

ISO 14937:2009

Sterilization of health care products - General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices

ISO 14937:2009 specifies general requirements for the characterization of a sterilizing agent and for the development, validation and routine monitoring and control of a sterilization process for medical devices.

It applies to sterilization processes in which microorganisms are inactivated by physical and/or chemical means and is intended to be applied by process developers, manufacturers of sterilization equipment, manufacturers of medical devices to be sterilized, and organizations responsible for sterilizing medical devices.

ISO 14937:2009 specifies the elements of a Quality Management System which are necessary to assure the appropriate characterization of the sterilizing agent, development, validation and routine monitoring and control of a sterilization process.

ISO 20857:2010

Sterilization of health care products - Dry heat - Requirements for the development, validation and routine control of a sterilization process for medical devices

ISO 20857:2010 specifies requirements for the development, validation and routine control of a dry heat sterilization process for medical devices.

Although ISO 20857:2010 primarily addresses dry heat sterilization, it also specifies requirements and provides guidance in relation to depyrogenation processes using dry heat.

ISO 25424:2018

Sterilization of health care products - Low temperature steam and formaldehyde - Requirements for development, validation and routine control of a sterilization process for medical devices

1.1 Inclusions

1.1.1 This document specifies requirements for the development, validation and routine control of a low temperature steam and formaldehyde (LTSF) sterilization process for medical devices using a mixture of low temperature steam and formaldehyde as sterilizing agent and which operates below ambient pressure.

NOTE Although the scope of this document is limited to medical devices, it specifies requirements and provides guidance that can be applicable to other products and equipment.

1.1.2 This document is intended to be applied by process developers, manufacturers of sterilization equipment, manufacturers of medical devices to be sterilized and the organizations with responsibility for sterilizing medical devices (see ISO 14937:2009, Table E.1).

1.2 Exclusions

1.2.1 This document does not specify requirements for the development, validation and routine control of a process for inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease. Specific recommendations have been produced in particular countries for the processing of materials potentially contaminated with these agents.

NOTE See ISO 22442‑1, ISO 22442‑2 and ISO 22442‑3.

1.2.2 This document does not specify requirements for designating a medical device as "STERILE". Such requirements are given in EN 556‑1.

1.2.3 This document does not specify a quality management system for the control of all stages of production of medical devices.

NOTE It is not a requirement of this document to have a complete quality management system during manufacture or reprocessing, but those elements of such a system that are required are normatively referenced at appropriate places in the text. Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production or reprocessing of medical devices including the sterilization process. Further guidance is given in E.4 of ISO 14937:2009.

1.2.4 This document does not specify requirements for occupational safety associated with the design and operation of LTSF sterilization facilities.

NOTE 1 Safety requirements for sterilizers are specified in IEC 61010‑2‑040.

NOTE 2 Attention is also drawn to the existence in some countries of regulations stipulating safety requirements.

1.2.5 This document does not cover analytical methods for determining levels or residues of formaldehyde and/or its reaction products.

NOTE 1 Attention is drawn to EN 14180.

NOTE 2 Attention is drawn to the possible existence in some countries of statutory regulations specifying limits for the level of formaldehyde residues on medical devices and products.

1.2.6 This document does not cover preparatory measures that might be necessary before sterilization such as cleaning, disinfection and packing.

NOTE For reprocessable medical devices, the manufacturer(s) of these devices can supply information on the preparatory measures (see ISO 17664).

ISO 11607 - Terminally Sterilized Medical Devices Package

ISO 11607 - Terminally Sterilized Medical Devices Package.

The ISO 11607 - Terminally Sterilized Medical Devices Package provides the requirements and test methods for packaging intended to maintain the sterility of terminally sterilized medical devices until the point of use. It also provides the requirements for the development and validation of processes including forming, sealing, and assembly of packaging used for sterilized medical devices. This package includes ISO 11607-1:2019, ISO 11607-2:2019. It has been reduced 15% off of the individual list price.

ISO 11607-1:2019

Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems

This document specifies requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.

It is applicable to industry, to health care facilities, and to wherever medical devices are placed in sterile barrier systems and sterilized.

It does not cover all requirements for sterile barrier systems and packaging systems for medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.

It does not describe a quality assurance system for control of all stages of manufacture.

It does not apply to packaging materials and/or systems used to contain a contaminated medical device during transportation of the item to the site of reprocessing or disposal.

ISO 11607-2:2019

Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes

This document specifies requirements for the development and validation of processes for packaging medical devices that are terminally sterilized. These processes include forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems.

It is applicable to industry, to health care facilities, and to wherever medical devices are packaged and sterilized.

It does not cover all requirements for packaging medical devices that are manufactured aseptically. Additional requirements can be necessary for drug/device combinations.

ISO/TS 16775:2021

Packaging for terminally sterilized medical devices - Guidance on the application of ISO 11607-1 and ISO 11607-2

This document provides guidance for the application of the requirements contained in ISO 11607-1 and ISO 11607-2. It does not add to, or otherwise change, the requirements of ISO 11607-1 and ISO 11607-2. This is an informative document, not normative. It does not include requirements to be used as basis of regulatory inspection or certification assessment activities. The guidance can be used to better understand the requirements of ISO 11607-1 and ISO 11607-2 and illustrates the variety of methods and approaches available for meeting the requirements of those International Standards. It is not required that this document be used to demonstrate conformity with them. Guidance is given for evaluation, selection and use of packaging materials, preformed sterile barrier systems, sterile barrier systems and packaging systems. Guidance on validation requirements for forming, sealing and assembly processes is also given. This document provides information for both healthcare facilities and the medical devices industry for terminally sterilized medical devices. This document does not provide guidance for applications of packaging materials and systems after their opening. In the use of packaging for other purposes such as a “sterile field” or transport of contaminated items, other regulatory standards will apply.

AAMI TIR52:2014 (R2017)

Environmental monitoring for terminally sterilized healthcare products

This TIR assists in establishing an environmental monitoring program that is meaningful, manageable and defendable, and provides guidance to avoid adverse environmental conditions during the manufacture of terminally sterilized healthcare products. 

ISO 11737 - Sterilization of Medical Devices Package

ISO 11737 - Sterilization Of Medical Devices Package (Save 19% off List Prices)

The ISO 11737 - Sterilization of Medical Devices Package provides the requirements to estimate and determine the population of viable microorganisms on a product and to test the sterility of a product that has been treated with a sterilizing agent. The determination of bioburden can be used to validate the sterilization process, monitor the manufacturing processes and assess the overall environmental monitoring programme. It has been reduced 25% off the list price. This package includes:
ISO 11737-1:2018
ISO 11737-2:2019

ISO 11737-1:2018

Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on products

ISO 11737-1:2018 specifies requirements and provides guidance on the enumeration and microbial characterization of the population of viable microorganisms on or in a health care product, component, raw material or package.

NOTE 1 The nature and extent of microbial characterization is dependent on the intended use of bioburden data.

NOTE 2 See Annex A for guidance on Clauses 1 to 9.

ISO 11737-1:2018 does not apply to the enumeration or identification of viral, prion or protozoan contaminants. This includes the removal and detection of the causative agents of spongiform encephalopathies, such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease.

NOTE 3 Guidance on inactivating viruses and prions can be found in ISO 22442‑3, ICH Q5A(R1) and ISO 13022.

ISO 11737-1:2018 does not apply to the microbiological monitoring of the environment in which health care products are manufactured.

ISO 11737-2:2019

Sterilization of health care products - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process

1.1 This document specifies the general criteria for tests of sterility on medical devices that have been exposed to a treatment with the sterilizing agent which has been reduced relative to that anticipated to be used in routine sterilization processing. These tests are intended to be performed when defining, validating or maintaining a sterilization process.

1.2 This document is not applicable to:

a) sterility testing for routine release of product that has been subjected to a sterilization process,

b) performing a test for sterility (see 3.12),

NOTE 1 The performance of a) or b) is not a requirement of ISO 11135, ISO 11137-1, ISO 11137-2, ISO 14160, ISO 14937, ISO 17665-1 or ISO 20857.

c) test of sterility or test for sterility for demonstration of product shelf life, stability and/or package integrity, and

d) culturing of biological indicators or inoculated products.

NOTE 2 Guidance on culturing biological indicators is included in ISO 11138-7.

ISO 17665-1:2006

Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices

ISO 17665-1:2006 specifies requirements for the development, validation and routine control of a moist heat sterilization process for medical devices.

Moist heat sterilization processes covered by ISO 17665-1:2006 include but are not limited to:

  • saturated steam venting systems;
  • saturated steam active air removal systems;
  • air steam mixtures;
  • water spray;
  • water immersion.

ISO/TS 17665-2:2009

Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1

ISO/TS 17665-2:2009 provides general guidance on the development, validation and routine control of moist heat sterilization processes and is intended to explain the requirements set forth in ISO 17665-1. The guidance given in this Technical Specification is provided to promote good practice related to moist heat sterilization processes and to assist those developing and validating a moist heat sterilization process according to ISO 17665-1.

ISO/TS 17665-3:2013

Sterilization of health care products - Moist heat - Part 3: Guidance on the designation of a medical device to a product family and processing category for steam sterilization

ISO/TS 17665-3:2013 provides guidance about the attributes of a medical device to be considered by the user when assigning a medical device to a product family for the purpose of identifying and aligning it to a processing category for a specific moist heat sterilization process.

PDA TR 30-2012

Parametric Release of Pharmaceuticals and Medical Device Products Terminally Sterilized by Moist Heat

This PDA technical report is intended to provide a single-source set of recommendations on developing a parametric release program that include demonstrated best practices that should be considered for development of a moist heat sterilization program. The report is built on the foundation of moist heat sterilization science presented in PDA Technical Report No. 1, which may be referred to for greater detail on the science of sterilization, sterilization cycle development and validation (2).

ISO 11040-4:2015

Prefilled syringes - Part 4: Glass barrels for injectables and sterilized subassembled syringes ready for filling

ISO 11040-4:2015 applies to

  • tubing-glass barrels (single-chamber design) for injection preparations, and
  • sterilized subassembled syringes ready for filling.

It specifies materials, dimensions, quality, and performance requirements, as well as relevant test methods.

ISO 11040-4:2015 also specifies those components that are part of the sterilized subassembled syringe ready for filling.

Glass barrels and sterilized subassembled syringes ready for filling in accordance with ISO 11040-4:2015 are intended for single use only.

Components to complete the subassembled syringe, such as plunger and rod, are not specified in ISO 11040‑4:2015.

ISO 11040-6:2019

Prefilled syringes - Part 6: Plastic barrels for injectables and sterilized subassembled syringes ready for filling

This document specifies materials, dimensions, quality, and performance requirements, as well as test methods for polymer barrels and sterilized subassembled syringes ready for filling, intended for single use only.

This document also specifies those components that are part of the sterilized subassembled syringe ready for filling.

Polymer barrels and sterilized subassembled syringes ready for filling in accordance with this document are intended for single use only.

Components to complete the subassembled syringe, such as plunger and rod, are not specified in this document.

Prefilled syringes can be produced on dedicated and specifically designed processing equipment such as inline moulding and filling. This document does not apply but can be used also for such dedicated prefilled syringes.

ISO 11040-7:2015

Prefilled syringes - Part 7: Packaging systems for sterilized subassembled syringes ready for filling

ISO 11040-7:2015 specifies the packaging system that is used to deliver sterilized subassembled syringes ready for filling in tubs and nests.

Downstream processes (processes after filling such as in house/outside transport, reprocessing) can result in specific requirements on the packaging system used to deliver sterilized subassembled syringes ready for filling. However, these requirements are not within the scope of ISO 11040-7:2015.