Most recent

ASTM F2902-16e1

Standard Guide for Assessment of Absorbable Polymeric Implants


1.1 This guide describes general guidelines for the chemical, physical, mechanical, biocompatibility, and preclinical assessments of implantable synthetic polymeric absorbable devices. This guide also describes evaluation methods that are potentially useful and should be considered when assessing absorbable implants or implant components.

1.2 The described evaluations may assist a manufacturer in establishing the safety and effectiveness of an absorbable implant device. This listing of assessment methods may also be utilized to assist in establishing substantial equivalence to an existing commercially marketed device. However, these polymeric material-oriented guidelines do not necessarily reflect the total needs for any particular implant application (for example, orthopedic, cardiovascular, sutures, and dermal fillers), which may require additional and potentially essential application-specific evaluations.

1.3 This guide is intended to cover all forms of absorbable polymeric components and devices, including solid (for example, injection-molded) and porous (for example, fibrous) forms. This guide is also intended to cover devices fabricated from amorphous and/or semi-crystalline absorbable polymer systems.

1.4 This guide has been generated with principal emphasis on the evaluation of devices formed from synthetic polymers that degrade in vivo primarily through hydrolysis (for example, α-hydroxy-polyesters). Evaluation methods suggested herein may or may not be applicable to implants formed from materials that, upon implantation, are substantially degraded through other mechanisms (for example, enzymatically induced degradation).


CONTENT PROVIDER
ASTM International [astm]

Others Also Bought
Implants for surgery - Homopolymers, copolymers and blends on poly(lactide) - In vitro degradation testing
Standard Guide for Characterization and Testing of Biomaterial Scaffolds Used in Regenerative Medicine and Tissue-Engineered ...
Biological evaluation of medical devices - Part 13: Identification and quantification of degradation products from ...
Included in Packages
This standard is not included in any packages.
Document History
Revises:
Amendments & Corrections
We have no amendments or corrections for this standard.