Historical

ASTM E694-99

Standard Specification for Laboratory Glass Volumetric Apparatus


This specification covers general requirements common to glass volumetric apparatus. Specific dimensions and tolerances for applicable instruments are given in other specifications as cited throughout this specification. Glass must conform to specifications E438 and be calibrated in accordance with Practice E542. 1.1.1 Class A -Each instrument shall be marked with the letter to signify compliance with applicable construction and accuracy requirements. Instruments may be marked with an identification marker (serial number) at the option of the manufacturer. 1.1.2 Class B -General purpose instruments are of the same basic design as Class A. However, volumetric tolerances for Class B instruments shall be within twice the specified range allowed for Class A unless otherwise specified.


CONTENT PROVIDER
ASTM International [astm]

Others Also Bought
Standard Specification for Laboratory Glass Graduated Burets
Standard Specification for Laboratory Glass Graduated Cylinders
Standard Specification for Laboratory Glass Volumetric Flasks
Document History
Revised By:
Included in Packages
This standard is not included in any packages.
Amendments & Corrections
We have no amendments or corrections for this standard.