Standard Test Method for Dynamic Impingement Between Femoral and Acetabular Hip Components

1.1 This test method covers a procedure to simulate dynamic impingement between femoral and acetabular components in a hip replacement; the subsequent qualitative assessment of damage modes (as outlined in 8.2); and, if necessary, quantitative assessment of changes in modular component attachment strength.

1.2 This test method can be used to evaluate impingement between femoral components and the following: single-piece, modular, semi-constrained, bipolar, constrained, or dual mobility acetabular components, manufactured from polymeric, metallic, or ceramic materials.

1.3 The values stated in SI units are regarded as the standard.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

CONTENT PROVIDER
ASTM International [astm]

Others Also Bought
Standard Specification and Test Methods for Metallic Medical Bone Screws
Standard Practice for Cyclic Fatigue Testing of Metal Tibial Tray Components of Total Knee Joint Replacements ...
Standard Guide for Evaluating Modular Hip and Knee Joint Components
Included in Packages
This standard is not included in any packages.
Document History
Revises:
Amendments & Corrections
We have no amendments or corrections for this standard.