Needle-based injection systems for medical use. Requirements and test methods Automated functions (British Standard)

This document specifies requirements and test methods for automated functions in needle-based injection systems with automated functions (NIS-AUTO).

 

General requirements are provided for all automated functions. In addition, specific requirements are provided for the following automated functions:

 

a) medicinal product preparation (e.g. reconstitution);

 

b) needle preparation;

 

c) needle hiding;

 

d) priming;

 

e) dose setting;

 

f) needle insertion;

 

g) injection depth control;

 

h) injection of the medicinal product;

 

i) recording of device functions;

 

NOTE  This document does not cover remote communication from the NIS-AUTO (pertains to wired and wireless communication transfer from the NIS auto).

 

j) disabling the NIS-AUTO;

 

k) needle retraction;

 

l) needle shielding;

 

m) needle removal.

 