Historical

ISO 14243-3:2004

Implants for surgery - Wear of total knee-joint prostheses - Part 3: Loading and displacement parameters for wear-testing machines with displacement control and corresponding environmental conditions for test


ISO 14243-3:2004 specifies relative movement between articulating components, the pattern of the applied force, speed and duration of testing, sample configuration and test environment to be used for the wear testing of total knee-joint prostheses in wear-testing machines having axial load control, flexion/extension angular motion control, AP displacement control and tibial rotation control.

The kinematics of ISO 14243-3:2004 may not be applicable to knee designs with a high degree of constraint, which could result in damage to the articulating components in the early stages of the test that would not be representative of clinical service.


CONTENT PROVIDER
International Organization for Standardization [iso]

Others Also Bought
Implants for surgery - Wear of total knee-joint prostheses - Part 1: Loading and displacement parameters for wear-testing ...
Implants for surgery -- Wear of total knee-joint prostheses -- Part 2: Methods of measurement
Implants for surgery - Wear of total hip-joint prostheses - Part 1: Loading and displacement parameters for wear-testing ...
Document History
Revised By:
Included in Packages
This standard is not included in any packages.
Amendments & Corrections
Corrected By: