Historical

ISO 7206-12:2016

Implants for surgery - Partial and total hip joint prostheses - Part 12: Deformation test method for acetabular shells

ISO 7206-12:2016 specifies a test method for determining short-term deformation of a press-fit acetabular component for total hip joint replacement under specific laboratory conditions. It also defines the conditions of testing so that the important parameters that affect the components are taken into account and it describes how the specimen is set up for testing. Furthermore, this part of ISO 7206 specifies the test parameters of press-fit acetabular components tested in accordance with this part of ISO 7206.

The described method is intended to be used to evaluate the comparison of various designs and materials used for acetabular components in total hip joint replacement when tested under similar conditions.

The loading of the acetabular components in vivo will, in general, differ from the loading defined in this test method. The results obtained here cannot be used to directly predict in vivo performance.

ISO 7206-12:2016 does not cover methods of examining the test specimen.

CONTENT PROVIDER
International Organization for Standardization [iso]

Others Also Bought
Implants for surgery - Partial and total hip joint prostheses - Part 4: Determination of endurance properties and ...
Implants for surgery - Partial and total hip-joint prostheses - Part 10: Determination of resistance to static ...
Implants for surgery - Partial and total hip joint prostheses - Part 6: Endurance properties testing and performance ...
Included in Packages
This standard is not included in any packages.
Document History
Revised By:
Amendments & Corrections
We have no amendments or corrections for this standard.