ASTM C1771-13

Standard Test Method for Determination of Boron, Silicon, and Technetium in Hydrolyzed Uranium Hexafluoride by Inductively Coupled Plasma—Mass Spectrometer After Removal of Uranium by Solid Phase Extraction

1.1 This test method covers the determination of boron, silicon and technetium in hydrolyzed uranium hexafluoride (UF6) by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) after separation of the uranium by solid phase extraction.

1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Some specific hazards statements are given in Section 7 on Hazards.


Content Provider
ASTM International [astm]


Document History
Revised By:
Included in Packages
This standard is not included in any packages.
Amendments & Corrections
We have no amendments or corrections for this standard.
ANSI Logo

As the voice of the U.S. standards and conformity assessment system, the American National Standards Institute (ANSI) empowers its members and constituents to strengthen the U.S. marketplace position in the global economy while helping to assure the safety and health of consumers and the protection of the environment.